Cubemass DCI MODBUS RS485 Technical data
Endress+Hauser 87
Example of maximum measured error
A0011691
Fig. 42: Max. measured error in % o.r. (example: Cubemass DCI, DN 1)
Flow values (examples)
Basis for calculations → ä 88.
Repeatability Basis for calculations → ä 88.
o.r. = of reading
Mass flow and volume flow (liquids)
• ±0.05% o.r. (mass flow)
• ±0.05% o.r. (volume flow)
Density (liquids)
• ±0.0005 g/cc
1 g/cc = 1 kg/l
Temperature
±0.25 °C ± 0.0025 · T °C
(±0.5 °F ± 0.0015 · (T – 32) °F)
T = Fluid temperature
Influence of medium
temperature
When there is a difference between the temperature for zero point adjustment and the process
temperature, the typical measured error is ±0.0002% of the full scale value / °C (±0.0001% of the
full scale value / °F).
lb/min0
0.8
0.2
1.00.4
0.6
[%]
kg/h
0
0.5
0.2
1.0
0.1
0.3
0.4
0.6
0.7
0.8
0.9
0
255101520
Flow rate Maximum measured error
[kg/h] [lb/min] [% o.r.]
0.1 0.0037 0.8
0.7 0.0257 0.114
2.5 0.0919 0.1
15 0.5513 0.1
o.r. = of reading