Cubemass DCI MODBUS RS485 Technical data
Endress+Hauser 85
10.1.5 Power supply
Electrical connections → ä 19
Supply voltage 85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
Cable entries Power supply and signal cables (inputs/outputs):
• Cable entry M20 × 1.5 (8 to 12 mm / 0.31 to 0.47")
• Threads for cable entries, ½" NPT, G ½"
Connecting cable for remote version:
• Cable entry M20 × 1.5 (8 to 12 mm / 0.31 to 0.47")
• Threads for cable entries, ½" NPT, G ½"
Cable specifications Remote version→ ä 21
Power consumption AC: < 15 VA (including sensor)
DC: < 15 W (including sensor)
Switch-on current
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure Lasting min. 1 power cycle:
• EEPROM or HistoROM T-DAT saves measuring system data if power supply fails.
• HistoROM/S-DAT: exchangeable data storage chip which stores the data of the sensor (nominal
diameter, serial number, calibration factor, zero point etc.)
Potential equalization No measures necessary.
For explosion-protected equipment → separate Ex-documentation supplied