Cubemass DCI MODBUS RS485 Commissioning
Endress+Hauser 51
6.3.3 Quick Setup "Gas measurement"
The measuring device is primarily designed for measuring liquid flow. The measurement of gases is
also possible.
!
Note!
Before carrying out the Quick Setup "Gas measurement", the Quick Setup "Commissioning" has
to be executed. ä 46
Only mass and Corrected volume flow can be measured and output with the gas measurement
mode. Note that direct density and/or volume measurement is not possible!
If corrected volume flow (e.g. in Nm
3
/h) is to be measured and output instead of the mass flow
(e.g. in kg/h), change the setting for the CORRECTED VOLUME CALCULATION function to
"FIXED REFERENCE DENSITY" in the "Commissioning" Quick Setup menu.
Corrected volume flow can be assigned as follows:
to a display line,
to the current output,
to the pulse/frequency output.