Cubemass DCI MODBUS RS485 Technical data
Endress+Hauser 83
10 Technical data
10.1 Technical data at a glance
10.1.1 Applications
→ ä 5
10.1.2 Function and system design
Measuring principle Mass flow measurement by the Coriolis principle
Measuring system → ä 7
10.1.3 Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring
tube to register a phase shift in the oscillation)
• Fluid density (proportional to the resonance frequency of the measuring tube)
• Fluid temperature (via temperature sensors)/ (not suitable for custody transfer measurement)
Measuring range
Measuring ranges for liquids
Operable flow range Over 1000 : 1
Flow rates above the preset full scale value do not overload the amplifier, i.e. the totalized flow
values are registered correctly.
Input signal Status input (auxiliary input)
U = 3 to 30 V DC, R
i
= 5 kΩ, galvanically isolated.
Configurable for: totalizer reset, positive zero return, error message reset, start zero point
adjustment.
DN Range for full scale values (liquids) g
min(F)
to g
max(F)
[mm] [inch] [kg/h] [lb/min]
1 1/24" 0 to 20 0 to 0.75
2 1/12" 0 to 100 0 to 3.70
4 1/8" 0 to 450 0 to 16.50
6 ¼" 0 to 1000 0 to 37.00