Cubemass DCI MODBUS RS485 Commissioning
Endress+Hauser 53
!
Note!
The parameters described in the table can be found in the "MODBUS RS485" group of the "BASIC
FUNCTION" block in the function matrix
(È "Description of Device Parameters" manual).
6304 PARITY Selection depends on the "Data transfer mode" function:
NONE; EVEN; UNEVEN
• Available in the ASCII transfer mode → even or uneven parity bit
(EVEN, UNEVEN).
• Available in the RTU transfer mode → no parity bit (NONE) or even
or uneven parity bit (EVEN, UNEVEN).
Factory setting: EVEN
6305 BYTE ORDER Select the byte transmission sequence for the Integer, Float and String
data types:
0 - 1 - 2 - 3
3 - 2 - 1 - 0
2 - 3 - 0 - 1
1 - 0 - 3 - 2
Factory setting: 1 - 0 - 3 - 2
!
Note!
The transmission sequence must suit the MODBUS master.
6306 DELAY TELE. REPLY For entering a delay time after which the measuring device replies to
the request telegram of the MODBUS master. This allows
communication to be adapted to slow MODBUS masters:
0 to 100 ms
Factory setting: 10 ms
Back to the HOME position:
→ Press and hold down Esc key X for longer than three seconds or
Repeatedly press and release → Esc keys X = Exit the function matrix step by step
Quick Setup "Communication"