Operation Cubemass DCI MODBUS RS485
40 Endress+Hauser
Access to data via MODBUS
The MODBUS master uses the register addresses 5051 to 5081 to access the data area of the auto-
scan buffer. This data area contains the values of the device parameters defined in the scan list.
For example, if the register 2007 was entered for mass flow in the scan list by means of the SCAN
LIST REG. 1 function, the master can read out the current measured value of the mass flow in
register 5051.
12 5012 SCAN LIST REG. 12
13 5013 SCAN LIST REG. 13
14 5014 SCAN LIST REG. 14
15 5015 SCAN LIST REG. 15
16 5016 SCAN LIST REG. 16
Data area
Parameter value/Measured values
Access via MODBUS
register address
Data type * Access**
Value of scan list entry No. 1 → 5051 Integer / Float Read/Write
Value of scan list entry No. 2 → 5053 Integer / Float Read/Write
Value of scan list entry No. 3 → 5055 Integer / Float Read/Write
Value of scan list entry No. 4 → 5057 Integer / Float Read/Write
Value of scan list entry No. 5 → 5059 Integer / Float Read/Write
Value of scan list entry No. 6 → 5061 Integer / Float Read/Write
Value of scan list entry No. 7 → 5063 Integer / Float Read/Write
Value of scan list entry No. 8 → 5065 Integer / Float Read/Write
Value of scan list entry No. 9 → 5067 Integer / Float Read/Write
Value of scan list entry No. 10 → 5069 Integer / Float Read/Write
Value of scan list entry No. 11 → 5071 Integer / Float Read/Write
Value of scan list entry No. 12 → 5073 Integer / Float Read/Write
Value of scan list entry No. 13 → 5075 Integer / Float Read/Write
Value of scan list entry No. 14 → 5077 Integer / Float Read/Write
Value of scan list entry No. 15 → 5079 Integer / Float Read/Write
Value of scan list entry No. 16 → 5081 Integer / Float Read/Write
* The data type depends on the device parameter entered in the scan list
** The data access depends on the device parameter entered in the scan list. If the device parameter entered supports read
and write access, the parameter can also be accessed by means of the data area.
Scan list
No. MODBUS configuration
Register address
(data type = Integer)
Configuration via
local operation / configuration program
(BASIC FUNCTION → MODBUS RS485 →)