What to do if Endress+Hauser Gammapilot M FMG60 show error W103?
- TTracy PhelpsSep 4, 2025
If your Endress+Hauser Measuring Instruments display error W103, wait for completion of the initializing procedure.
What to do if Endress+Hauser Gammapilot M FMG60 show error W103?
If your Endress+Hauser Measuring Instruments display error W103, wait for completion of the initializing procedure.
What to do if Endress+Hauser Measuring Instruments show error A116?
If your Endress+Hauser Measuring Instruments display error A116, repeat the download.
What to do if Endress+Hauser Gammapilot M FMG60 Measuring Instruments show error A121?
If your Endress+Hauser Measuring Instruments display error A121 due to an electronics defect, first try switching the device off and then on again. If the error persists after restarting, contact Endress+Hauser service or consider replacing the transmitter.
What to do if density not calibrated on Endress+Hauser Gammapilot M FMG60 Measuring Instruments?
To resolve a 'density not calibrated' error with Endress+Hauser Measuring Instruments: * Verify that at least one calibration point has been entered and activated within the calibration function group. If not, enter and activate the necessary calibration points. * Ensure that the reference pulse rate is not larger than 232. If it is, re-perform the density calibration within the calibration function group.
What to do if Endress+Hauser Gammapilot M FMG60 Measuring Instruments show error A113?
If your Endress+Hauser Measuring Instruments display error A113 due to an electronics defect, first try switching the device off and then on again. If the error persists after restarting, contact Endress+Hauser service or consider replacing the transmitter.
What to do if Endress+Hauser Gammapilot M FMG60 show error A111?
If your Endress+Hauser Measuring Instruments display error A111 due to an electronics defect, first try switching the device off and then on again. If the error persists after restarting, contact Endress+Hauser service or consider replacing the transmitter.
What to do if Endress+Hauser Gammapilot M FMG60 Measuring Instruments show error A165?
If your Endress+Hauser Measuring Instruments display error A165 due to an electronics defect, first try switching the instrument off and then on again. If the error persists after restarting, contact Endress+Hauser service or consider replacing the transmitter.
What happens when Endress+Hauser Measuring Instruments issues a warning error?
When a warning error occurs, the Endress+Hauser Measuring Instruments device continues measurement, and an error message is displayed, alternating with the measured value.
What to do if Endress+Hauser Gammapilot M FMG60 Measuring Instruments shows the last error?
The last error is displayed in the "diagnostics" function group in the "previous error" function. The displayed item can be deleted in the "delete last error" function.
What happens when Endress+Hauser Gammapilot M FMG60 issues an alarm error?
When an alarm error occurs in Endress+Hauser Measuring Instruments, the output signal assumes a value set using the "output on alarm" function. This value can be MAX (110%, 22mA), MIN (-10%, 3.8mA), Hold (last value is on hold), or a user-specific value. Additionally, an error message is displayed.
Measurement Principle | Gamma radiation absorption |
---|---|
Housing Material | Aluminum or stainless steel |
Protection Class | IP66/67 |
Radiation Source | Cesium-137 (Cs-137) or Cobalt-60 (Co-60) |
Accuracy | ±1% of measuring range |
Output Signal | 4 to 20 mA HART |
Power Supply | 10.5 to 45 V DC |
Ambient Temperature | -20 to +60 °C |
Communication | HART |
Certifications | ATEX, IECEx |
Application | Density, level, or concentration measurement in liquids, solids, suspensions or slurries |
Process Temperature | -40 to +80 °C |
Covers safety regulations and procedures for radiation protection.
Details installation for density/concentration measurement.
Shows terminal assignments for compartments A and B.
Provides step-by-step wiring instructions for compartment A.
Explains how to operate the device using the display module.
Describes the FOUNDATION Fieldbus interface and network architecture.
Covers parameters for device calibration and configuration.
Provides a checklist for Fieldbus commissioning.
Provides an overview of calibration procedures.
Guides through the basic setup steps for the device.
Explains calibration for level and limit detection.
Lists system error signals and types.
Provides a table of error codes and their remedies.
Provides a menu tree for level measurement functions.