Technical data Proline Promass 83 PROFIBUS DP/PA
158 Endress+Hauser
10.1.9 Operating conditions: Process
Medium temperature range Sensor:
Promass F, A, H:
–50...+200 °C
Promass F (high-temperature version):
–50...+350 °C
Promass M, I:
–50...+150 °C
Promass E:
–40...+125 °C
Seals:
Promass F, E, H, I:
No internal seals
Promass M:
Viton –15...+200 °C; EPDM –40...+160 °C; silicon –60...+200 °C; Kalrez –20...+275 °C;
FEP sheathed (not for gas applications): –60...+200 °C
Promass A
(only for mounting sets with threaded connections)
Viton –15...200 °C; EPDM –40...+160 °C; silicon –60...+200 °C; Kalrez –20...+275 °C
Limiting medium pressure
range (rated pressure)
The material load diagrams (pressure-temperature diagrams) for the process connections are to be
found in the following documents:
• Technical Information Promass 80/83 F, M (TI 053D/06/en)
• Technical Information Promass 80/83 E (TI 061D/06/en)
• Technical Information Promass 80/83 A (TI 054D/06/en)
• Technical Information Promass 80/83 H, I (TI 052D/06/en)
Pressure ranges of secondary containment:
Promass F:
DN 8...50: 40 bar or 600 psi; DN 80: 25 bar or 375 psi;
DN 100...150: 16 bar or 250 psi; DN 250: 10 bar or 150 psi
Promass M:
100 bar or 1500 psi
Promass E:
No secondary containment
Promass A:
25 bar or 375 psi