Proline Promass 83 PROFIBUS DP/PA Technical data
Endress+Hauser 149
Recommended full scale values
See Page 159 ff. (“Limiting flow”)
Operable flow range Greater than 1000 : 1. Flows above the preset full scale value do not overload the amplifier, i.e.
totalizer values are registered correctly.
Input signal Status input (auxiliary input):
U = 3...30 V DC, R
i
= 3 k, galvanically isolated.
Switch level: ±3...±30 VDC, independent of polarity
Configurable for: totalizer reset, positive zero return, error message reset, start zero point
adjustment, batching start/stop (optional), batching totalizer return (optional)
10.1.4 PROFIBUS DP output
Output signal Current output
Active/passive selectable, galvanically isolated, time constant selectable (0.05...100 s),
Full scale value adjustable, temperature coefficient: typ. 0.005% o.r./°C, resolution: 0.5 µA
Active: 0/4...20 mA, R
L
max. 700
Passive: 4...20 mA; supply voltage V
S
18...30 V DC; R
i
150
Pulse / frequency output:
Active/passive selectable, galvanically isolated
Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), R
L
> 100
Passive: open collector, 30 V DC, 250 mA
Frequency output: full scale frequency 2...10000 Hz (f
max
= 12500 Hz), on/off ratio 1:1, pulse
width max. 2 s
Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05…2000 ms)
PROFIBUS DP interface:
PROFIBUS DP in accordance with IEC 61158, galvanically isolated
Profile Version 3.0
Data transmission rate: 9.6 kBaud...12 MBaud
Automatic identification of data transmission rate
Signal coding: NRZ code
Bus address can be configured via miniature switches or via the local display (optional)
PROFIBUS PA interface:
PROFIBUS PA in accordance with IEC 61158 (MBP), galvanically isolated
Data transmission rate: 31.25 kBaud
Current consumption: 11 mA
Permitted supply voltage: 9...32 V
Bus connection with integrated reverse polarity protection
Error current FDE (Fault Disconnection Electronic): 0 mA
Signal coding: Manchester II
Bus address can be configured via miniature switches or via local display (optional)
Signal on alarm Current output
Failsafe mode selectable (for example NAMUR recommendation NE 43)
Pulse/frequency output
Failsafe mode selectable