Commissioning Proline Promass 83 PROFIBUS DP/PA
78 Endress+Hauser
6.4.2 Advanced diagnostic functions
Changes to the measuring system, e.g. coating build-up or corrosion and abrasion on the measuring
tubes can be detected at an early stage by means of the optional software package “Advanced
Diagnostics” (F-Chip, accessories Page 117). Normally, these influences reduce the measuring
accuracy of the system or may lead to serious system errors.
By means of the diagnostic functions it is now possible to record various process and device
parameters during operation, e.g. mass flow, density/reference density, temperature values,
measuring tube damping etc.
By analysing the trend of these measured values, deviations of the measuring system from a
“reference status” can be detected in good time and corrective measures can be taken.
Reference values as the basis for trend analysis
Reference values of the parameters in question must always be recorded for trend analysis. These
reference values are determined under reproducible, constant conditions. Such reference values are
initially recorded during calibration at the factory and saved in the measuring device.
Reference data can also be ascertained under customer-specific process conditions, e.g. during
commissioning or at certain process stages (cleaning cycles, etc.).
Reference values are recorded and saved in the measuring system always by means of the device
function “REFERENCE CONDITION USER” (7401).
"
Caution!
It is not possible to analyse the trend of process/device parameters without reference values!
Reference values can only be determined under constant, non-changing process conditions.
Methods of ascertaining data
Process and device parameters can be recorded in two different ways which you can define in the
function ACQUISITION MODE (7410):
PERIODICAL option: Measuring device acquires data periodically. Enter the desired time interval
by means of the function ACQUISITION PERIOD (7411).
SINGLE SHOT option: The user acquires the data manually at different, free selectable periods.
Ensure that the process conditions always correspond to the reference status when data is being
recorded. It is only in this way that deviations from the reference status can be clearly determined.
!
Note!
The last ten entries are retained in chronological order in the measuring system.
The “history” of such values can be called up via various functions:
Diagnosis parameters Data saved (per parameter)
Mass flow
Density
Reference density
Temperature
Measuring tube damping
Sensor symmetry
Reference value “REFERENCE VALUE” function
Lowest measured value “MINIMUM VALUE” function
Highest measured value “MAXIMUM VALUE” function
List of the last ten measured values “HISTORY” function
Deviation measured/reference value “ACTUAL DEVIATION” function
More detailed information can be found in the “Description of Device Functions” Manual.