Commissioning Proline Promass 83 PROFIBUS DP/PA
68 Endress+Hauser
6.3.4 Quick Setup “Gas Measurement”
The measuring device is not only suitable for measuring liquid flow. Direct mass measurement
based on the Coriolis principle is also possible for measuring the flow rate of gases. For this purpose,
the relevant device functions can be configured either by means of a configuration program or the
local display.
!
Note!
Only mass and Corrected volume flow can be measured and output with the gas measurement
mode. Note that direct density and/or volume measurement is not possible!
The flow ranges and measuring accuracy that apply to gas measurement are not the same as those
for liquids.
Performing the “Gas Measurement” Quick Setup
The “Gas Measurement” Quick Setup menu guides you systematically through the setup procedure
for all the device functions that have to be configured for operation with gas.
The “Gas Measurement” Quick Setup can be called up in the function matrix by means of two
different functions:
Via the QS GAS MEASUREMENT (1004) function Fig. 40 or
Via the QS COMMISSIONING (1002) function Page 59
a0002601-en
Fig. 40: Quick Setup “Gas Measurement”
Recommended settings are found on the following page.
+
+
E
1004B
6400
6402
6403
E
Esc
E
+
-
XXX.XXX.XX
Setup
Gas measurement
Quick
Setup
Assign
Low flow cut off
On value
Low flow cut off
Off value
Low flow cut off
HOME