Proline Promass 83 PROFIBUS DP/PA Commissioning
Endress+Hauser 67
Pulsating Flow” Quick Setup menu”
HOME position → F → MEASURED VARIABLE (A)
MEASURED VARIABLE → O → QUICK SETUP (B)
QUICK SETUP → N → QS PULS. FLOW (1003)
Function No. Function name Selection with ( P )
1003 QS PULS. FLOW YES
After F is pressed by way of confirmation, the
Quick Setup menu calls up all the subsequent
functions in succession.
▼
Basic configuration
2002 DISPLAY DAMPING 1 s
3002 TOTALIZER MODE (DAA) BALANCE (Totalizer 1)
3002 TOTALIZER MODE (DAB) BALANCE (Totalizer 2)
3002 TOTALIZER MODE (DAC) BALANCE (Totalizer 3)
Signal type for “CURRENT OUTPUT 1...n”
4004 MEASURING MODE PULS. FLOW
4005 TIME CONSTANT 1 s
Signal type for “FREQ./PULSE OUTPUT 1...n” (for FREQUENCY operating mode)
4206 MEASURING MODE PULS. FLOW
4208 TIME CONSTANT 0 s
Signal type for “FREQ./PULSE OUTPUT 1...n” (for PULSE operating mode)
4225 MEASURING MODE PULS. FLOW
Other settings
8005 ALARM DELAY 0 s
6400 ASSIGN LOW FLOW CUTOFF MASS FLOW
6402 ON-VALUE LOW FLOW CUT OFF Setting depends on diameter:
DN 1 = 0.02 [kg/h] resp. [l/h]
DN 2 = 0.10 [kg/h] resp. [l/h]
DN 4 = 0.45 [kg/h] resp. [l/h]
DN 8 = 2.0 [kg/h] resp. [l/h]
DN 15 = 6.5 [kg/h] resp. [l/h]
DN 15* = 18 [kg/h] resp. [l/h]
DN 25 = 18 [kg/h] resp. [l/h]
DN 25* = 45 [kg/h] resp. [l/h]
DN 40 = 45 [kg/h] resp. [l/h]
DN 40* = 70 [kg/h] resp. [l/h]
DN 50 = 70 [kg/h] resp. [l/h]
DN 80 = 180 [kg/h] resp. [l/h]
DN 100 = 350 [kg/h] resp. [l/h]
DN 150 = 650 [kg/h] resp. [l/h]
DN 250 = 1800 [kg/h] resp. [l/h]
* DN 15, 25, 40 “FB” = Full bore versions of
Promass I
6403 OFF-VALUE LOW FLOW CUTOFF 50%
6404 PRESSURE SHOCK SUPPRESSION 0 s
▼
Back to the HOME position:
→ Press and hold down Esc key X for longer than three seconds or
→ Repeatedly press and release Esc key X → Exit the function matrix step by step