Operation Proline Promass 83 PROFIBUS DP/PA
54 Endress+Hauser
5.6.4 Current output configuration
The current output is configured as “active” or “passive” by means of various jumpers on the
I/O board or the current submodule.
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages.
Make sure that the power supply is switched off before you remove the cover of the electronics
compartment.
1. Switch off power supply
2. Remove the I/O board Page 137 ff.
3. Set the jumpers Fig. 32
"
Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in the diagram.
Incorrectly set jumpers can cause overcurrents that would destroy either the measuring
device or external devices connected to it.
Note that the position of the current submodule on the I/O board can vary, depending on
the version ordered, and that the terminal assignment in the connection compartment of the
transmitter varies accordingly.
4. Installation of the I/O board is the reverse of the removal procedure.
F06-x3MBxxxx-16-xx-06-xx-001
Fig. 32: Configuring current outputs with the aid of jumpers (I/O board)
1 Current output
1.1 Active current output (default)
1.2 Passive current output
1.1
1.2
+
+
INPUT/OUTPUT 4
INPUT/OUTPUT 3
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1
2
3
4
W
E
N
O
1

Table of Contents

Other manuals for Endress+Hauser Proline Promass 83

Related product manuals