Proline Promass 83 PROFIBUS DP/PA Commissioning
Endress+Hauser 59
6.3 Quick Setup
In the case of measuring devices without a local display, the individual parameters and functions
must be configured via the configuration program, e.g. Fieldcare or ToF Tool - Fieldtool Package. If
the measuring device is equipped with a local display, all the important device parameters for
standard operation can be configured quickly and easily by means of the “Commissioning” Quick
Setup menu.
Quick Setup “Commissioning”, see next section
Quick Setup “Batching”, Page 65 ff.
Quick Setup “Pulsating flow”, Page 65 ff.
Quick Setup “Gas measurement”, Page 68 ff.
Quick Setup “Communication”, Page 70 ff.
Saving the settings and parameters, T-DAT administration
6.3.1 Quick SetupCommissioning
If the measuring device is equipped with a local display, all the important device parameters for
standard operation can be configured quickly and easily by means of the “Commissioning” Quick
Setup menu.
In the case of measuring devices without a local display, the individual parameters and functions
must be configured via a configuration program, (ToF Tool – Fieldtool Package, Fieldcare).
!
Note!
The display returns to the cell SETUP COMMISSIONING (1002) if you press the ESC key combination during
parameter interrogation. The stored parameters remain valid.
The “Commissioning” Quick Setup must be carried out before one of the Quick Setups explained below is run.
m Only units not yet configured in the current Setup are offered for selection in each cycle. The unit for mass, volume
and corrected volume is derived from the corresponding flow unit.
n The “YES” option remains visible until all the units have been configured.
“NO” is the only option displayed when no further units are available.
o This prompt only appears if a current output and/or frequency output is available. Only the outputs not yet
configured in the current Setup are offered for selection in each cycle.
p The “YES” option remains visible until all the outputs have been parameterized.
“NO” is the only option displayed when no further outputs are available.
q The “automatic parameterisation of the display” option contains the following basic settings/factory settings
YES Main line = mass flow
Additional line = totalizer 1
Information line = operating/system condition
NO The existing (selected) settings remain.
r The QUICK SETUP BATCHING is only available when the optional software package BATCHING is installed and
PROFIBUS DP is present.
s The QUICK SETUP PULSATING FLOW is only available when a current output and/or frequency output is
available.