Commissioning Proline Promass 83 PROFIBUS DP/PA
70 Endress+Hauser
6.3.5 Quick Setup “Communication”
To establish cyclic data transfer, various arrangements between the PROFIBUS master and slave are
required which have to be taken into consideration when configuring various functions. These
functions can be configured quickly and easily by means of the “Communication” Quick Setup. The
table explains the parameter configuration options in more detail.
a0002600-en
Fig. 41: Communication Quick Setup.
+
+
E
+
E
B
Quick SetupQuick Setup
6101
6140
6141
1006
Esc
E
+
-
XXX.XXX.XX
HOME-POSITION
Bus address
Selection
GSD
Set unit to bus
Setup
Communication
Communication” Quick Setup.
HOME position → F → MEASURED VARIABLE (A)
MEASURED VARIABLE → O → QUICK SETUP (B)
QUICK SETUP → N → QUICK SETUP COMMUNICATION (1006)
Function No. Function name Setting to be selected ( P )
(to next function with F )
1006 QUICK SETUP
COMMUNICATION
YES → After F is pressed by way of confirmation, the Quick
Setup menu calls up all the subsequent functions in succession.
6101 BUS ADDRESS Enter the device address (permitted address range: 1...125)
Factory setting: 126
6140 SELECTION GSD Select the operating mode (GSD file) which should be used for
cyclic communication with the PROFIBUS master.
Options:
MANUFACT. SPEC. → The Promass 83 measuring device is
operated in the manufacturer-specific mode.
MANUFACT V2.0 → The Promass 83 measuring device is used as
the replacement for the previous Promass 63 model (compatibility
mode).
GSD PROFILE → The Promass 83 measuring device is operated in
the Profile mode.
Factory setting: MANUFACT. SPEC.