A3 Size Color Flat Bed Scanner GT-30000 Revision A
DISASSEMBLY & ASSEMBLY Disassembly Procedures 49
4.2.2.5  Glass Frame Assembly Removal
1. Remove the mechanism cover. ("Chapter 4.2.1.3 Mechanism Cover 
Removal".)
2. Remove the 12 screws (No.5), and then remove the glass frame assembly. 
("Figure 4-15. Glass Frame Assembly Removal".)
Figure 4-15.  Glass Frame Assembly Removal
4.2.2.5.1  Glass Frame Assembly Installation
After the glass frame assembly has been removed, be sure to follow the 
procedure given below when re-installing the glass frame assembly.
1. Provisionally place the glass frame assembly on top of the scanner 
mechanism in the position specified.
2. Determine the installation position while referring to "Figure 4-16. Glass 
Frame Assembly Positioning".
3. Once the glass frame assembly installation position has been determined, 
secure it in place by tightening the 12 screws.
C A U T I O N
!
Place the scanner on a stable, level surface when 
removing the glass frame assembly.
!
When re-installing the glass frame assembly, be sure 
to check the installation position thoroughly while 
referring to "Chapter 4.2.2.5.1 Glass Frame Assembly 
Installation". (If the glass frame assembly is not 
installed in the correct position, the document 
scanning start position will not be correct.)
A D J U S T M E N T
R E Q U I R E D
After re-installing or replacing the glass frame assembly, 
be sure to adjust the home position of the scanner, by 
using the special software. (See "Chapter 5.3.4 Home 
Position Adjustment".)
C H E C K
P O I N T
When installing the glass frame assembly, the glass 
fixing plate should be installed correctly. (See "Figure 
4-16. Glass Frame Assembly Positioning".)
Screws
ScrewsScrews
Screws(No.
(No.(No.
(No.
5
)
))
)
Glass fixing plate
Screws
ScrewsScrews
Screws(No.
(No.(No.
(No.
5
)
))
)