7 • Maintenance and servicing
66 / 84
80116-938_V24392
2023-03
AUTO START thermoSEAL
(for an MF socket)
1. Plug a 20100-152 cable into the first MF socket.
2. Set the test specimen to:
BiClamp, bipolar– thermoSEAL – Effect 1 –AUTO START
3. Hold down the S1 button on the VIO test box until the test specimen
is activated. Activation must begin after the test delay that has been
set.
4. Press the S2 button. The test specimen must terminate the activation.
5. Unplug cable.
6. Repeat test steps 1 to 5 with each MF socket.