7 • Maintenance and servicing
68 / 84
80116-938_V24392
2023-03
1. When prompted to do so, select the type of return electrode: e.g. split
return electrode (the type of return electrode selected in this window
only remains set until the next time the system is started. After that,
the test specimen resumes operation using the preferred type of
return electrode as stored).
2. On the VIO, press the <NESSY> button .
The monitoring window for the return electrode is opened that shows
the resistance.
3. Determine and document the measured value. The tolerance range is
123 to 143 ohms.