7 | Technical data FISCHER Mess- und Regeltechnik GmbH
96/108 BA_EN_FT90
Measuring range Measurement error
[%]
Tc-zero
[%/10K]
Tc span
[%/10K]
Typ. Max. Typ. Max. Typ. Max.
-4 to +4 mbar -400 to +400 Pa 0.5 1.0 0.1 0.2 0.05 0.1
-6 to +6 mbar -600 to +600 Pa 0.5 0.75 0.1 0.15 0.05 0.1
-10 to +10 mbar -1 to +1 kPa 0.25 0.5 0.05 0.1 0.05 0.1
Sensor type B
Measuring range Measurement error
[%]
Tc-zero
[%/10K]
Tc span
[%/10K]
Typ. Max. Typ. Max. Typ. Max.
0 to 4 mbar 0 to 400 Pa 0.5 1.0 0.15 0.3 0.05 0.1
0 to 6 mbar 0 to 600 Pa 0.5 0.75 0.15 0.25 0.05 0.1
0 to 10 mbar 0 to 1 kPa 0.25 0.5 0.1 0.2 0.05 0.1
0 to 16 mbar 0 to 1.6 kPa 0.25 0.5 0.15 0.3 0.05 0.1
0 to 25 mbar 0 to 2.5 kPa 0.25 0.5 0.15 0.25 0.05 0.1
0 to 40 mbar 0 to 4 kPa 0.25 0.5 0.1 0.2 0.05 0.1
0 to 60 mbar 0 to 6 kPa 0.25 0.5 0.1 0.2 0.05 0.1
0 to 100 mbar 0 to 10 kPa 0.25 0.5 0.1 0.15 0.05 0.1
0 to 160 mbar 0 to 16 kPa 0.25 0.5 0.05 0.1 0.05 0.1
0 to 250 mbar 0 to 25 kPa 0.25 0.5 0.05 0.1 0.05 0.1
-2.5 to +2.5 mbar -250 to +250 Pa 0.5 1.0 0.15 0.3 0.05 0.1
-4 to +4 mbar -400 to +400 Pa 0.5 1.0 0.1 0.2 0.05 0.1
-6 to +6 mbar -600 to +600 Pa 0.5 0.75 0.1 0.15 0.05 0.1
-10 to +10 mbar -1 to +1 kPa 0.25 0.5 0.05 0.1 0.05 0.1
-16 to +16 mbar -1.6 to +1.6 kPa 0.25 0.5 0.1 0.2 0.05 0.1
-25 to +25 mbar -2.5 to +2.5 kPa 0.25 0.5 0.1 0.15 0.05 0.1
-40 to +40 mbar -4 to +4 kPa 0.25 0.5 0.05 0.1 0.05 0.1
-60 to +60 mbar -6 to +6 kPa 0.25 0.5 0.05 0.1 0.05 0.1
-100 to +100 mbar -10 to +10 kPa 0.25 0.5 0.05 0.1 0.05 0.1
-160 to +160 mbar -16 to +16 kPa 0.25 0.5 0.05 0.1 0.05 0.1
-250 to +250 mbar -25 to +25 kPa 0.25 0.5 0.05 0.1 0.05 0.1
7.5 Digital interfaces
USB interface
USB On The Go 2.0
Data rate 12 Mbit/s (Full Speed)
Port Micro USB type B
Communication Host/Device mode