Operator‘s Manual FERITSCOPE
®
FMP30 Page 101
Measuring
The readings and the block results of a single block can be printed during
the evaluation of this particular block by pressing the PRINT key.
8.1.1 ‘Documenting the Block Result With a Printer’, beginning on
Page 121
Fig. 7-8 Printout of the block results of a single block (example)
Explanation to Fig. 7-5 up to Fig. 7-8:
FISCHER FERIT-
SCOPE FMP30
Instrument type:
2008-07-24 Current date
Application No. Number of the Application
Block No. Number of the open measurement block
n Consecutive number of the reading
fe
Measured value of the
-ferrite content
LSL/USL Upper / lower specification limit
* Reading is within the tolerance limits
<< / >> Reading violates the tolerance limits
fe. Mean value of all readings obtained for this step of the cal-
ibration.
s Standard deviation of the readings obtained for this step of
the calibration.
C. O. V. Coefficient of variation
FISCHER FERITSCOPE FMP30 24.07.08
Appl.No. 0 -- Block result --
Block No.: 3 24.07.08 12:09
Mean value fe. = 10.15 +/- 0.13%
Std. dev. s = 0.11% n = 5
C. O. V. = 1.06%
lowest meas.= 10.07%
highest meas. = 10.33%
4 values < LSL ( 10.20% )= 80.00%
0 values > USL ( 10.80% )= 0.00%