Page 176 Operator‘s Manual FERITSCOPE
®
FMP30
Instrument Settings - Service Function Menu
ENTER
Select the desired “delay” by pressing the
arrow keys
or and confirm the selec-
tion with ENTER.
[0 ms]: no pause after external start
[100 ms]: 100 ms pause after external start
to
[2500 ms]: 2500 ms pause after external
start
[Page : CAL]: Press CAL to leaf through
the pages.
Use DEL at any time to cancel the setting
procedure.
The combination “measurement effect off” and “external start off” is
not permissible because it would not allow for a measurement acqui-
sition. If this combination is set, the instrument will automatically reset
the parameters to “measurement effect on” and “external start off”.
Key sequ. /
Action