Page 196 Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Unable to generate the
final result when mak-
ing measurements with
fixed block sizes if the
first block of the evaluat-
ed Application is not
closed.
Take additional mea-
surements until the first
block is concluded and
re-create the final result.
118
Unable to perform cor-
rective calibration with
the set
-ferrite content
(e.g., because [Base 2:
0] was set).
Set the rated
-ferrite
content values for the
used calibration stan-
dard and continue with
the corrective calibra-
tion.
64
No probe connected. Connect a probe. 25
Probe not connected
correctly.
Connect probe correct-
ly.
25
Probe defective. Use properly function-
ing probe.
25
When overwriting an
Application, it is possi-
ble to retain the mea-
surement data only of
the connected probe us-
es the same measure-
ment method as the
probe that was used to
set up the Application.
Delete the measure-
ment data when over-
writing the Application
or set up a new Appli-
cation with the connect-
ed probe.
102
Unable to store the next
reading in this block be-
cause the block is full
(applies to matrix mode
only).
Use BLOCK-RES and
arrow keys to change
the block.
118
Malfunction/Message Cause Correction
start
on
Page
E 017
Not enough
measurements !
E 021
Entered Standard
invalid !
E 022
Missing probe !
E 023
Meas. method
different !
E 024
Result block
overflow !