Operator‘s Manual FERITSCOPE
®
FMP30 Page 195
Malfunctions and Messages
Cause: Calibration stan-
dards used that did not
exhibit the required
-
ferrite content or are de-
fective.
Repeat the master cali-
bration with correct and
good calibration stan-
dards.
70
Instrument-internal er-
ror: Unable to compute
the coefficients of the
master characteristic
(original master charac-
teristic will be retained).
Repeat the master cali-
bration.
If this error occurs re-
peatedly: Inform cus-
tomer service.
70
Unable to store master
characteristic.
Cause: Probe not
plugged in and tight-
ened properly.
Plug probe in and tight-
en properly; repeat the
master calibration.
25
Cause: Probe defective. Connect a properly
functioning probe and
if applicable repeat the
master calibration.
25
70
The stored
-ferrite con-
tents do not correspond
to the measured stan-
dards.
Cause: Calibration stan-
dards were measured
out of sequence during
the master calibration
(standard 1 mixed up
with standard 2) and
-
ferrite content was not
adjusted corresponding-
ly using the arrow keys.
Repeat the master cali-
bration and measure
the standards in the cor-
rect sequence.
70
Malfunction/Message Cause Correction
start
on
Page
E 014
Unable to calc.
parameters !
E 015
Unable to store
cal. in probe !
E 016
Std. and meas.
not matching !