Page 194 Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Unable to complete cor-
rective calibration.
Cause: Measurement
carried out incorrectly.
Repeat the corrective
calibration and perform
the calibration measure-
ment correctly (e.g., do
not allow the probe to
hover over the standard
before or after the mea-
surement; observe the
minimum lift-off dis-
tance!)
64
Cause: Calibration stan-
dards used that did not
exhibit the required
-
ferrite content or are de-
fective.
Repeat the corrective
calibration with correct
and good calibration
standards.
64
Calibration standards
were measured out of
sequence during the cor-
rective calibration (stan-
dard 1 mixed up with
standard 2) and
-fer-
rite content was not ad-
justed correspondingly
using the arrow keys.
Repeat the corrective
calibration and mea-
sure the standards in
the correct sequence.
64
Unable to complete the
master calibration.
Cause: Measurement
carried out incorrectly.
Repeat the master cali-
bration and perform the
calibration measure-
ment correctly (e.g., do
not allow the probe to
hover over the standard
before or after the mea-
surement; observe the
minimum lift-off dis-
tance!)
70
Malfunction/Message Cause Correction
start
on
Page
E 011
Measurements
out
E 012
Invalid std.
sequence !
E 013
Countrate out
of interval !