Operator‘s Manual FERITSCOPE
®
FMP30 Page 193
Malfunctions and Messages
Cause: ENTER was
pressed during the mea-
surement in the “free-run-
ning” display mode
while the probe was lift-
ed off.
Place the probe on the
specimen and then
press ENTER to trig-
ger the measurement
acquisition.
105
Outlier measurement
recognized during nor-
malization of calibra-
tion.
Cause: Measurement on
calibration standard car-
ried out incorrectly.
Repeat the calibration
step and perform the
calibration measure-
ment correctly (e.g., do
not allow the probe to
hover over the standard
before or after the mea-
surement; observe the
minimum lift-off dis-
tance!)
34
Cause: Measurement
carried out on wrong
calibration standard.
Repeat the calibration
step and make the cali-
bration measurement
on a correct standard.
64
Connected probe not
suitable or not support-
ed (measurement meth-
od of the connected
probe does not fit the in-
strument).
Connect a suitable
probe.
25
Probe defective. Connect a properly
functioning probe.
25
Malfunction/Message Cause Correction
start
on
Page
E 007
Measurements
invalid !
E 010
Meas. method
not supported !