Page 62 Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
6.2.1 Normalization Procedure
Key sequ. /
Action
Detail of the display Explanation
ZERO Use ZERO to start the normalization of the
open Application.
ZERO appears and remains on the display
while the normalization is performed.
[s]: Standard deviation
[n]: Number of measurements
[Base material (Fe]: The measurements
should be made on the base material of the
calibration standard set.
[Cancel: ENTER]: Using ENTER cancels
the normalization.
Base
Make about 5 measurements at different lo-
cations of the base.
The mean value of all readings obtained for
the normalization will be displayed.
[Delete: DEL]: Use DEL to delete the last
measurement, 2x DEL to delete all read-
ings obtained for the normalization.
[OK: ENTER]: Using ENTER ends the nor-
malization.
ENTER A confirmation indicating that the normal-
ization has been carried out successfully
appears.
Pressing ENTER confirms the message.
PRINT: A print form of the normalization
will be printed if a printer is connected and
switched on ( Page 63).