Operator‘s Manual FERITSCOPE
®
FMP30 Page 69
Normalization, Calibration and Master Calibration
6.3.4 Documenting the Corrective Calibration with a Printer
Fig. 6-1 Print form of a corrective calibration (example)
Explanations to Fig. 6-1:
FISCHER FERIT-
SCOPE FMP30
Instrument type:
2008-07-23 Current date
CALIBRATION Corrective calibration is being documented
Date Time Date and time of the corrective calibration
Appl. No. Number of the Application
Probe Probe that has been used to perform the calibration
Fe Applied measurement method
Fe:
Confirmed rated value for the
-ferrite content of the cali-
bration standards.
fe. Mean value of all readings obtained for this step of the
calibration.
s Standard deviation of the readings obtained for this step
of the calibration.
Measurements on
Base
Measurements on
Calibration stan-
dard # 1
Measurements on
Calibration stan-
dard # 2
FISCHER FERITSCOPE FMP30 23.07.08
CALIBRATION 23.07.08 12:43
Appl.No.: 3 Probe:FGAB1.3Fe
Fe
fe.= 104.2 Fe% s= 1.671 Fe%
Fe: 0.680 Fe%
fe.= 0.646 Fe% s= 0.010 Fe%
Fe: 3.08 Fe%
fe.= 3.054 Fe% s= 0.038 Fe%
Fe: 10.90 Fe%
fe.= 10.35 Fe% s= 0.014 Fe%
Measurements on
Calibration stan-
dard # 3