Page 68 Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
6.3.3 Deleting a Corrective Calibration
When deleting the corrective calibration, only the corrective calibra-
tion of the open Application will be deleted. The corrective calibra-
tions of the other Applications are retained. However, if the linking
mode is enabled ( on the display), the corrective calibrations
of all Applications that are linked with the open Application will be
deleted.
Key sequ. /
Action
Detail of the display Explanation
CAL Use CAL to start the corrective calibration
of the open Application.
[Delete cal: DEL]: Using DEL deletes the
corrective calibration.
[Cancel: CAL]: Using CAL cancels the cor-
rective calibration (the stored corrective cal-
ibration is retained unchanged).
DEL [Yes: DEL]: The corrective calibration will
be deleted.
[No: ENTER]: Cancels the delete proce-
dure.
DEL The corrective calibration will be deleted.
The instrument is again ready to make mea-
surements.