Operator‘s Manual FERITSCOPE
®
FMP30 Page 67
Normalization, Calibration and Master Calibration
To verify a correct calibration, perform several reference measurements
on certified calibration standards. The corrective calibration must be
repeated if the mean values obtained from several single readings are
not within the guaranteed error limit stated on the calibration stan-
dards, o if the deviations are unacceptable again, a master calibration
must be performed.
ENTER A confirmation indicating that the corrective
calibration has been carried out successful-
ly appears.
Pressing ENTER confirms the message.
PRINT: A print form of the corrective cali-
bration will be printed if a printer is con-
nected and switched on ( Page 69).
ENTER The new characteristic will be computed au-
tomatically and stored. The instrument is
again ready to make measurements.
( 2.1 ‘Display’, beginning on Page 8.)
Key sequ. /
Action
Detail of the display Explanation