Page 66 Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
ENTER
[Entry:
]: Use the arrow keys to set the
rated
-ferrite content value of the calibra-
tion standard.
The rated value can be set faster if a mea-
surement is performed on a calibration stan-
dard that is within the tolerance range, and
then the rated value is corrected using the
arrow keys.
[CAL-rat. 1: 0.680]: Display of the set rat-
ed value for the
-ferrite content of the cali-
bration standard (Example: 0.680 Fe%)
[Cancel: ENTER]: Using ENTER cancels
the calibration procedure
Calibration
standard
Make about 5 measurements at different lo-
cations of the calibration standard.
Displayed is the mean value of all measure-
ments performed for this step.
[Delete: DEL]: Use DEL to delete the last
measurement, 2x DEL to delete all read-
ings obtained for the calibration.
[OK: ENTER]: Use ENTER to end the cur-
rent calibration step.
ENTER If a corrective calibration is desired with 2
or 3 standards, proceed in the same man-
ner with calibration standards 2 and 3.
Otherwise: Use ENTER to end the correc-
tive calibration.
The new characteristic will be computed au-
tomatically and stored.
The instrument is again ready to make mea-
surements.
Key sequ. /
Action
Detail of the display Explanation