Operator‘s Manual FERITSCOPE
®
FMP30 Page 73
Normalization, Calibration and Master Calibration
ENTER ZERO + CAL appear and remain on the
display while the master calibration is per-
formed.
[s]: Standard deviation
[n]: Number of measurements
[Base material (Fe)]: The measurements
(normalization) should be made on base.
[MCL Delete: DEL]: Use DEL to delete an
existing master calibration (appears only if
a master calibration has already been per-
formed). Use CAL if you do not wish to de-
lete the master calibration.
[Cancel: CAL]: Use CAL to cancel the
master calibration.
If you press DEL again after deleting the master calibration in order to
exit the menu, subsequent measurements will be made with the factory
master calibration - without a corrective calibration.
Key sequ. /
Action
Detail of the display Explanation