Page 74 Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
Key sequ. /
Action
Detail of the display Explanation
Base
Make about 5 measurements at different lo-
cations of the base.
The mean value of all readings obtained for
the normalization will be displayed.
[Base material (Fe)]: The measurements
should be made on base.
[Delete: DEL]: Use DEL to delete the last
measurement DEL, 2x DEL to delete all
readings obtained for the normalization.
[OK: ENTER]: Use ENTER to end and
store the normalization. The previous nor-
malization will be overwritten.
ENTER
[Entry:
]: Use the arrow keys to set the
rated
-ferrite content value of the calibra-
tion standard.
The rated value can be set faster if a mea-
surement is performed on the calibration
standard and then the rated value is correct-
ed using the arrow keys.
[CAL-rat. 1: 680]: Display of the set rated
value for the
-ferrite content of the calibra-
tion standard (Example: 0.680 Fe%)
[Cancel: ENTER]: Using ENTER cancels
the calibration procedure
Pressing FINAL-RES:
Determination of the Normalized Coun-
trate Xn of a Calibration Standard During a
Master Calibration Page 79.
Pressing BLOCK-RES:
Displaying Xn Ranges for Calibration Stan-
dards for the Master Calibration,
Page 76.