Operator‘s Manual FERITSCOPE
®
FMP30 Page 75
Normalization, Calibration and Master Calibration
Calibration
standard
Make about 5 measurements at different lo-
cations of the calibration standard.
[Delete: DEL]: Use DEL to delete the last
measurement DEL, 2x DEL to delete all
readings obtained for the normalization.
[OK: ENTER]: Use ENTER to end and
store the calibration step.
or Use the arrow keys or to set the rated
-ferrite content value of the calibration
standard.
The rated value can be set faster if a mea-
surement is performed on the calibration
standard and then the rated value is correct-
ed using the arrow keys
or .
This step is not required if the rated value
for the -ferrite content coincides with the
stored (displayed) value.
[OK: ENTER]: Using ENTER ends the cur-
rent calibration step.
ENTER Proceed with calibration standard #2 in the
same as described for calibration standard
#1. The same for all subsequent calibration
steps.
ENTER After the last calibration step has been con-
cluded using ENTER, the new calibration
curve will be computed automatically and
stored, and a confirmation indicating that
the master calibration has been carried out
successfully appears.
Pressing ENTER confirms the message.
PRINT: A print form of the master calibra-
tion will be printed if a printer is connected
and switched on ( Page 78).
Key sequ. /
Action
Detail of the display Explanation