EasyManuals Logo

Flintec DAD143 Series User Manual

Flintec DAD143 Series
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
DAD 143.x User Manual Profinet Rev.1.1.1 March 2023 Page 38 of 95
Zero point, gain and decimal point position were saved in the EEPROM; the calibration counter (TAC) is
increased automatically by +1.
9.2.2 Value
[CM1: Index 0x2300 Sub 0x07] [CM2: Sub 0x0E] [CM3: Sub 0x0F]
point in multi range applications). Permitted values are from 1 to 999 999.
Master (PC / PLC) sends
Slave (DAD143.x) responds
Meaning
CM
M+050000
Request: CM = 50000 d
CE
E+00017 (example)
Request: TAC counter CE17
CE 17
OK
Calibration sequence active
CM 30000
OK
Setup: CM = 30000 d
This value will determine the -range
respective the point at which the output will change the measuring range / interval size.
Application
CM 1 = MAX 1
CM 2 = MAX 2
CM 3 = MAX 3
Single range
CM 1 = 1...999 999
CM 2 = 0
(means CM 2 not used)
CM 3 = 0
(means CM 3 not used)
Dual range or dual interval
( Command MR)
CM 1 = 1...MAX 1
CM 2 = MAX 1...999 999
Triple range or triple interval
Teilungen ( Befehl MR)
CM 1 = 1...MAX 1
CM 2 = MAX 1...MAX 2
CM 3 = MAX 2...999 999
It is necessary: MAX 1 < MAX 2 < MAX 3
Note: The range, in which a scale can be set to zero (SZ) or automatic zero tracking (ZT) is active, is +/- 2% of
CM value. But for none legal for trade applications you can change the behaviour with the settings of ZT (see
10.2.8) and/or ZR (see 10.2.10).
Factory default: CM1 = 999 999, CM 2 = 0, CM 3 = 0
9.2.3 CI Set Minimum Output Value
[Index 0x2300 Sub 0x08]
This command is used to set up the minimum output value. Permitted values are from 999 999 to 0.
Master (PC / PLC) sends
Slave (DAD143.x) responds
Meaning
CI
I 010009
Request: CI = 10009 d
CE
E+00017 (example)
Request: TAC counter CE17
CE 17
OK
Calibration sequence active
CI 100
OK
Setup: CI = 100 d
This value will -range.
Note: In bipolar applications (e.g. force- or torque measurements) this parameter defines the max. output
value for input signals with negative sign.
Factory default: CI = 010009
9.2.4 MR Set Multi-range / Multi-interval
[Index 0x2300 Sub 0x0D]
This command is only relevant, if CM 2 > 0 or CM 3 > 0. Is this the case, this command defines, if the
application is multi-range or multi-interval weighing. Permitted values are 0 (Multi-interval) or 1 (Multi-range).
Master (PC / PLC) sends
Slave (DAD143.x) responds
Meaning
MR
M+00000
Request: MR = 0 (Multi-interval)
CE
E+00017 (example)
Request: TAC counter CE17
CE 17
OK
Calibration sequence active
MR 1
OK
Setup: MR = 1 (Multi-range)
Note: Single range applications ignore this parameter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flintec DAD143 Series and is the answer not in the manual?

Flintec DAD143 Series Specifications

General IconGeneral
BrandFlintec
ModelDAD143 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals