DAD 143.x User Manual Profinet Rev.1.1.1 March 2023 Page 66 of 95
13.2.3 NG Network Gateway (Read only)
The NG command can show the actual default gateway address currently in use by the DAD 143.x.
PC sends
DAD 143.x responds
Meaning
NG
G:000.000.000.000
Current default gateway is shown
Factory default: 0.0.0.0
Note: The NG setting can also be viewed from the front panel. Use front panel menu 8.6.
13.2.4 DN Device Name (Read only)
The DN command can show the actual name of station currently in use by the DAD 143.x
PC sends
DAD 143.x responds
Meaning
DN
N:dad143
Factory default: No name
13.2.5 MA MAC Addresses (Read only)
The DAD 143.x is assigned four consecutive MAC addresses during manufacture. These MAC addresses cannot
be changed by the user but they can be inspected using the MA command.
PC sends
DAD 143.x responds
Meaning
MA1
00-02-A2-50-4A-47
Show MAC address 1
MA2
00-02-A2-50-4A-48
Show MAC address 2
MA3
00-02-A2-50-4A-49
Show MAC address 3
MA4
00-02-A2-50-4A-4A
Show MAC address 4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flintec DAD143 Series and is the answer not in the manual?

Flintec DAD143 Series Specifications

General IconGeneral
BrandFlintec
ModelDAD143 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals