DAD 143.x User Manual Profinet Rev.1.1.1 March 2023 Page 43 of 95
Note:
Please note, the firmware type selection is locked in the same way as the calibration, that means it must be
unlocked with the command "CE n" before the firmware type can be set. Then the FT setting must be saved
with the command "CS".
9.3 Motion Detection Commands NR, NT
The motion detection facility provides a means of disabling certain functions whenever a condition of
-
steady for the period of time set by NT, during which it cannot fluctuate by more than NR increments. The
stable condition activates the relev
not stable.
9.3.1 NR -
[Index 0x2100 Sub 0x0A]
Permitted values are from 1 to 65535.
Slave (DAD143.x) responds
Factory default: NR = 1 [= ±1d]
9.3.2 NT -
[Index 0x2100 Sub 0x0B]
- iod NT to be considered
Permitted values are from 1 to 65535.
Slave (DAD143.x) responds
If the value of NT = 500 milliseconds, the output must not fluctuate more than NR increments within 500
Factory default: NT = 1000 [ms]
9.4 Filter Setting Commands FM, FL, PF, UR
A digital filter can be set which will eliminate most of the unwanted disturbances. The commands FM and FL
are used to define the digital filter settings, the command UR is used to define an averaging of up to 128
measurement values. Please note that these filters are positioned immediately after the A/D Converter and
therefore affect all aspects of the weighing operation.
9.4.1 FM Filter Mode
[Index 0x2100 Sub 0x09]
This command defines the filter mode. Choose the filter mode for your application.
Slave (DAD143.x) responds
Request: FM = 0 (IIR filter)
Setup: FM = 0 (IIR filter)