DAD 143.x User Manual Profinet Rev.1.1.1 March 2023 Page 11 of 95
3.7 LED Indicators Terminal 27
LED
Colour
State
Description
Sys
On
Protocol firmware running
On or flashing
Service/maintenance firmware running in various modes
BF
System Failure
Off
No error
Flashing, 1Hz
DCP signal service is initiated via the bus
On
System error
SF
Bus Failure
Off
No error
Flashing, 2Hz
No data exchange
On
No configuration or no physical link
Lnk1
On or flashing
Port 1 Ethernet link established
Act1
Flashing
Port 1 Ethernet Rx/Tx activity
Lnk2
On or flashing
Port 2 Ethernet link established
Act2
Flashing
Port 2 Ethernet Rx/Tx activity
Table of LED indicators for ProfiNet protocol
3.8 Seal Switch
Traceable Access Code (TAC) protected calibration commands
In case the seal switch is closed, the following commands or menus be proceeded:
- Calibration Zero
- Calibration Gain
- Calibration Absolute Zero
- Calibration Absolute Gain
- Calibration Minimum
- Calibration Maximum
- Zero Tracking
- Zero Range
- Display Step Size
- Decimal Point
- Calibration Save
- Factory Default
- Non Volatile Tare
- Non Volatile Zero
- Initial Zero @ power ON
- etc.
Activity Link of Port 1/2
Setup or changes of calibration can only be performed with an
open seal switch (26). Changes lead to get a new TAC value,
increased by +1.
Running a legal for trade application needs the jumper
connected to the switch pins and a seal. A broken seal shows
up changes of calibration, which are not allowed.
Protected commands see below.