Front Panel Operation
How to Set Output 4
4-27
nn084f.eps
8. Press the numeric keys and decimal point key to enter the desired current output
(maximum six numeric keys). For example, 234.567.
9. Press a multiplier key, if necessary. For example, press c.
10. Press A.
11. The Control Display now shows the amplitude of your voltage and current entries.
For example, 123.456 mV and 234.567 mA (below).
nn085f.eps
12. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. Then press the H key. For example,
1.1234 kHz.
13. The Control Display now shows your entries. For example, 123.456 mV and
234.567 mA at 1.1234 kHz (below).
nn086f.eps
14. Press E. The calibrator clears your entry from the Control Display and copies it
into the Output Display (below is typical).
nn087f.eps
15. Press O to activate the calibrator output. When changing power output levels, you
must reenter both voltage and current (in either order).
(Enter voltage or current and then a watts entry value using EA. The remaining
volts or current value is calculated and displayed.)
Three softkey labels appear on the Control Display: WAVE MENUS, I OUT (AUX or
20A terminals), and LCOMP (off or on). The Control Display also shows the real power
output for sine waves. Power out is computed as Power = Cosine Φ (Volts x Current)
where Φ is the phase difference between the volts and current waveforms. Cosine Φ is
also known as the Power Factor (PF).