5522A
Operators Manual
9-4
General Specifications
Voltage Function Specifications
DC Signal Square Wave Signal
[1]
Voltage Function
50 Ω Load 1 MΩ Load 50 Ω Load 1 MΩ Load
Amplitude Characteristics
Range
0 to ±6.599 V 0 to ±130 V
±1 mV to
±6.599 V p-p
±1 mV to
±130 V p-p
Resolution
Range
1 to 24.999 mV
25 to 109.99 mV
110 mV to 2.1999 V
2.2 to 10.999 V
11 to 130 V
Resolution
1 μV
10 μV
100 μV
1 mV
10 mV
Adjustment Range Continuously adjustable
1-Year Absolute Uncertainty,
tcal ±5 °C
±(0.25 % of output
+ 40 μV)
± 0.05 % of output
+ 40 μV)
±(0.25 % of output
+ 40 μV)
±(0.1 % of
output + 40 μV)
[2]
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range 10 Hz to 10 kHz
1-Year Absolute Uncertainty,
tcal ±5 °C
±(2.5 ppm of setting)
Typical aberration within 4 μs from
50 % of leading/trailing edge
<(0.5 % of output + 100 μV)
[1] Selectable positive or negative, zero referenced square wave.
[2] For square wave frequencies above 1 kHz, ± (0.25 % of output + 40 μV).
Edge Specifications
Edge Characteristics into 50 Ω Load
1-Year Absolute Uncertainty,
tcal ± 5 °C
Rise Time
≤300 ps
(+0 ps / -100 ps)
Amplitude Range (p-p) 4.5 mV to 2.75 V
±(2 % of output + 200 μV)
Resolution 4 digits
Adjustment Range
±10 % around each sequence value
(indicated below)
Sequence Values
5 mV, 10 mV, 25 mV, 50 mV, 60 mV, 80
mV, 100 mV, 200 mV, 250 mV, 300 mV,
500 mV, 600 mV, 1 V, 2.5 V
Frequency Range
[1]
900 Hz to 11 MHz
±(2.5 ppm of setting)
Typical Jitter, edge to trigger <5 ps (p-p)
within 2 ns from 50 % of rising edge <(3 % of output + 2 mV)
2 to 5 ns <(2 % of output + 2 mV)
5 to 15 ns <(1 % of output + 2 mV)
Leading Edge Aberrations
[2]
after 15 ns <(0.5 % of output + 2 mV)
Typical Duty Cycle 45 % to 55 %
Tunnel Diode Pulse Drive Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 to 100 V p-p.
[1] Above 2 MHz rise time specification <350 ps
[2] All edge aberration measurements made with Tektronix 11801 mainframe with SD26 input module.