5522A
Operators Manual
10-34
Parameters: 1 to 60 (seconds)
Example: TLIMIT 30
TLIMIT?
(IEEE-488, RS-232, Sequential)
Returns the programmed OPERATE time limit for the OVERLD mode signal.
Response: <Integer> Time limit in seconds.
TLIMIT_D
(IEEE-488, RS-232, Sequential)
Sets the default OPERATE time limit for the OVERLD mode signal.
Parameters: 1 to 60 (seconds)
Example: TLIMIT_D 15
TLIMIT_D?
(IEEE-488, RS-232, Sequential)
Returns the default overload time limit.
Response: <Integer> Default time limit in seconds.
Impedance/Capacitance Function Commands
ZERO_MEAS
(IEEE-488, RS-232, Sequential)
Sets the measurement offset to the capacitance value.
Parameters: (boolean) ON or OFF.
*TRG
(IEEE-488, RS-232, Sequential)
Triggers and returns a new impedance measurement value when used with the SC1100
option in MEAS Z mode. (See Chapter 6 for *TRG use in all cases except MEAS Z mode
with the SC1100 option.)
Responses: <measurement value>, OHM (input impedance value in ohms)
<measurement value>, F (input capacitance value in farads)
<measurement value>, NONE (no measurement is available)
Example: *TRG returns 1.00E+03,OHM (1 kΩ input impedance).
Note
You can also use the VAL? query to return an impedance measurement
value with the SC1100 option. VAL? returns the last measurement, whereas
*TRG gets a new measurement. Responses are the same as shown above for
the *TRG command. (See Chapter 6 for VAL? use with thermocouple
measurements.)