SC1100 Oscilloscope Calibration Option
General Specifications10
10-7
Pulse Generator Specifications
Pulse Generator Characteristics
Positive pulse into 50 Ω
Typical rise/fall times <1.5 ns
Available Amplitudes 2.5 V, 1 V, 250 mV, 100 mV, 25 mV, 10 mV
Pulse Width
Range
4 to 500 ns
[1]
Uncertainty (typical)
5 % ±2 ns
Pulse Period
Range 20 ms to 200 ns (50 Hz to 5 MHz)
Resolution 4 or 5 digits depending upon frequency and width
1-Year Absolute Uncertainty at Cardinal Points, tcal
± 5 °C
±2.5 ppm
[1] Pulse width not to exceed 40 % of period.
[2] Pulse width uncertainties for periods below 2 μs are not specified.
Trigger Signal Specifications (Pulse Function)
Pulse Period Division Ratio
Amplitude into 50 Ω (p-p)
Typical Rise Time
20 ms to 150 ns
off/1/10/100
≥ 1 V ≤2 ns
Trigger Signal Specifications (Time Marker Function)
Time Marker Period Division Ratio
Amplitude into 50 Ω (p-p)
Typical Rise Time
5 s to 35 ms
off/1
≥1 V ≤2 ns
34.9 ms to 750 ns
off/1/10/100
≥1 V ≤2 ns
749 to 7.5 ns
off/10/100
≥1 V ≤2 ns
7.4 to 2 ns
off/100
≥1 V ≤2 ns
Trigger Signal Specifications (Edge Function)
Edge Signal
Frequency
Division Ratio
Typical Amplitude into 50 Ω
(p-p)
Typical Rise Time Typical Lead Time
1 kHz to 10 MHz
off/1
≥1 V ≤2 ns
40 ns
Trigger Signal Specifications (Square Wave Voltage Function)
Voltage Function
Frequency
Division Ratio
Typical Amplitude into 50
Ω (p-p)
Typical Rise Time
Typical Lead
Time
10 Hz to 10 kHz
off/1
≥1 V ≤2 ns 2 μs
TV Trigger Signal Specifications
Trigger Signal Type Parameters
Field Formats Selectable NTSC, SECAM, PAL, PAL-M
Polarity Selectable inverted or uninverted video
Amplitude into 50 Ω (p-p) Adjustable 0 to 1.5 V p-p into 50 ohm load, (±7 % accuracy)
Line Marker Selectable Line Video Marker
Oscilloscope Input Resistance Measurement Specifications
Scope input selected
50 Ω 1 MΩ
Measurement Range
40 to 60 Ω 500 kΩ to 1.5 MΩ
Uncertainty 0.1 % 0.1 %