5522A
Operators Manual
10-44
Leveled Sinewave Verification: Frequency
Table 10-15. SC1100 Option Leveled Sinewave Verification: Frequency
Nominal Value
(V p-p)
Frequency Measured Value
(Hz)
Deviation (Hz) 1-Year Spec.
(Hz)
5.5 50 kHz 0.125
5.5 500 kHz 1.25
5.5 5 MHz 12.5
5.5 50 MHz 125
5.5 500 MHz 1250
3.5 1000 MHz 2500
Leveled Sinewave Verification: Harmonics
Table 10-16. SC1100 Option Leveled Sinewave Verification: Harmonics
Harmonic
Nominal
Value (V p-p)
Frequency
Measured
Value (dB)
Deviation
(dB)
1-Year Spec.
(dB)
2nd harmonic 0.0399 50 kHz -33
3rd+ harmonic 0.0399 50 kHz -38
2nd harmonic 0.099 50 kHz -33
3rd+ harmonic 0.099 50 kHz -38
2nd harmonic 0.399 50 kHz -33
3rd+ harmonic 0.399 50 kHz -38
2nd harmonic 1.2 50 kHz -33
3rd+ harmonic 1.2 50 kHz -38
2nd harmonic 5.5 50 kHz -33
3rd+ harmonic 5.5 50 kHz -38
2nd harmonic 5.5 100 kHz -33
3rd+ harmonic 5.5 100 kHz -38
2nd harmonic 5.5 200 kHz -33
3rd+ harmonic 5.5 200 kHz -38
2nd harmonic 5.5 400 kHz -33
3
rd
+ harmonic 5.5 400 kHz -38
2
nd
harmonic 5.5 800 kHz -33
3
rd
+ harmonic 5.5 800 kHz -38
2
nd
harmonic 5.5 1 MHz -33
3
rd
+ harmonic 5.5 1 MHz -38
2
nd
harmonic 5.5 2 MHz -33