EasyManua.ls Logo

Fluke 5522A - AC Voltage (Sine Wave)

Fluke 5522A
426 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5522A
Operators Manual
1-12
AC Voltage (Sine Wave)
Absolute Uncertainty,
tcal ±5 °C
±(ppm of output + μV)
Range Frequency
90 days 1 year
Resolution
Max
Burden
Max Distortion and
Noise
10 Hz to 5 MHz
Bandwidth
±(% output + floor)
Normal Output
10 Hz to 45 Hz 600 + 6 800 + 6
0.15 + 90 μV
45 Hz to 10 kHz 120 + 6 150 + 6
0.035 + 90 μV
10 kHz to 20 kHz 160 + 6 200 + 6
0.06 + 90 μV
20 kHz to 50 kHz 800 + 6 1000 + 6
0.15 + 90 μV
50 kHz to 100 kHz 3000 + 12 3500 + 12
0.25 + 90 μV
1.0 mV to
32.999 mV
100 kHz to 500 kHz 6000 + 50 8000 + 50
1 μV 65 Ω
0.3 + 90 μV
[1]
10 Hz to 45 Hz 250 + 8 300 + 8
0.15 + 90 μV
45 Hz to 10 kHz 140 + 8 145 + 8
0.035 + 90 μV
10 kHz to 20 kHz 150 + 8 160 + 8
0.06 + 90 μV
20 kHz to 50 kHz 300 + 8 350 + 8
0.15 + 90 μV
50 kHz to 100 kHz 600 + 32 800 + 32
0.20 + 90 μV
33 mV to
329.999 mV
100 kHz to 500 kHz 1600 + 70 2000 + 70
1 μV 65 Ω
0.20 + 90 μV
[1]
10 Hz to 45 Hz 250 + 50 300 + 50
0.15 + 200 μV
45 Hz to 10 kHz 140 + 60 150 + 60
0.035 + 200 μV
10 kHz to 20 kHz 160 + 60 190 + 60
0.06 + 200 μV
20 kHz to 50 kHz 250 + 50 300 + 50
0.15 + 200 μV
50 kHz to 100 kHz 550 + 125 700 + 125
0.20 + 200 μV
0.33 V to
3.29999 V
100 kHz to 500 kHz 2000 + 600 2400 + 600
10 μV
10 mA
0.20 + 200 μV
[1]
10 Hz to 45 Hz 250 + 650 300 + 650 0.15 + 2 mV
45 Hz to 10 kHz 125 + 600 150 + 600 0.035 + 2 mV
10 kHz to 20 kHz 220 + 600 240 + 600 0.08 + 2 mV
20 kHz to 50 kHz 300 + 600 350 + 600 0.2 + 2 mV
3.3 V to
32.9999 V
50 kHz to 100 kHz 750 + 1600 900 + 1600
100 μV
10 mA
0.5 + 2 mV
45 Hz to 1 kHz 150 + 2000 190 + 2000 0.15 + 10 mV
1 kHz to 10 kHz 160 + 6000 200 + 6000 0.05 +10 mV
10 kHz to 20 kHz 220 + 6000 250 + 6000 0.6 + 10 mV
20 kHz to 50 kHz 240 + 6000 300 + 6000 0.8 + 10 mV
33 V to
329.999 V
50 kHz to 100 kHz 1600 + 50000 2000 + 50000
1 mV
5 mA,
except
20 mA for
45 Hz to
65 Hz
1.0 + 10 mV
45 Hz to 1 kHz 250 + 10000 300 + 10000 0.15 + 30 mV
1 kHz to 5 kHz 200 + 10000 250 + 10000 0.07 + 30 mV
330 V to
1020 V
5 kHz to 10 kHz 250 + 10000 300 + 10000
10 mV
2 mA,
except 6 mA
for 45 Hz to
65 Hz
0.07 + 30 mV
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9 % of output + floor as shown.
Note
Remote sensing is not provided. Output resistance is <5 mΩ for outputs 0.33 V. The AUX output resistance is <1 Ω. The maximum
load capacitance is 500 pF, subject to the maximum burden current limits

Table of Contents

Related product manuals