Introduction and Specifications
Detailed Specifications 1
1-13
AC Voltage (Sine Wave) (cont.)
Absolute Uncertainty,
tcal ±5 °C
±(% of output + μV)
Range
Frequency
[1]
90 days 1 year
Resolution
Max
Burden
Max Distortion and
Noise
10 Hz to 5 MHz
Bandwidth
±(% output + floor)
AUX Output
10 Hz to 20 Hz 0.15 + 370 0.2 + 370
0.2 + 200 μV
20 Hz to 45 Hz 0.08 + 370 0.1 + 370
0.06 + 200 μV
45 Hz to 1 kHz 0.08 + 370 0.1 + 370
0.08 + 200 μV
1 kHz to 5 kHz 0.15 + 450 0.2 + 450
0.3 + 200 μV
5 kHz to 10 kHz 0.3 + 450 0.4 + 450
0.6 + 200 μV
10 mV to
329.999 mV
10 kHz to 30 kHz 4.0 + 900 5.0 + 900
1 μv
5 mA
1 + 200 μV
10 Hz to 20 Hz 0.15 + 450 0.2 + 450
0.2 + 200 μV
20 Hz to 45 Hz 0.08 + 450 0.1 + 450
0.06 + 200 μV
45 Hz to 1 kHz 0.07 + 450 0.09 + 450
0.08 + 200 μV
1 kHz to 5 kHz 0.15 + 1400 0.2 + 1400
0.3 + 200 μV
5 kHz to 10 kHz 0.3 + 1400 0.4 + 1400
0.6 + 200 μV
0.33 V to
3.29999 V
10 kHz to 30 kHz 4.0 + 2800 5.0 + 2800
10 μv
5 mA
1 + 200 μV
10 Hz to 20 Hz 0.15 + 450 0.2 + 450
0.2 + 200 μV
20 Hz to 45 Hz 0.08 + 450 0.1 + 450
0.06 + 200 μV
45 Hz to 1 kHz 0.07 + 450 0.09 + 450
0.08 + 200 μV
1 kHz to 5 kHz 0.15 + 1400 0.2 + 1400
0.3 + +200 μV
3.3 v to 5 v
5 kHz to 10 kHz 0.3 + 1400 0.4 + 1400
100 μv
5 mA
0.6 + 200 μV
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is <5 mΩ for outputs 0.33 V. The AUX output resistance is <1 Ω. The maximum
load capacitance is 500 pF, subject to the maximum burden current limits