10-38
Table 10-4. SC1100 Option AC Voltage Verification (cont.)
(1 MΩ output impedance unless noted)
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-year Spec.
(V p-p)
200 mV 10000 0.00054
2.2 V 100 0.00224
2.2 V 5000 0.00554
2.2 V 10000 0.00554
Table 10-5. SC1100 Option AC Voltage Verification at 50
Calibrator
Mainframe
Output
(1 kHz)
HP 3458
Range
Topline
Reading
Baseline
Reading
Peak-to-Peak
Peak-to-
Peak x
Correction
Tolerance
(±V)
1 mV 100 mV dc 0.000043
-1 mV 100 mV dc 0.000043
10 mV 100 mV dc 0.000065
-10 mV 100 mV dc 0.000065
25 mV 100 mV dc 0.000103
-25 mV 100 mV dc 0.000103
110 mV 100 mV dc 0.000315
-110 mV 100 mV dc 0.000315
500 mV 1 V dc 0.00129
-500 mV 1 V dc 0.00129
2.2 V 10 V dc 0.00554
-2.2 V 10 V dc 0.00554
6.6 V 10 V dc 0.01654
-6.6 V 10 V dc 0.01654
AC Voltage Frequency Verification
Table 10-6. SC1100 Option AC Voltage Frequency Verification
(1 MΩ output impedance unless noted)
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(Hz)
Deviation (Hz)
1-year Spec.
(Hz)
2.1 10 0.000025
2.1 100 0.00025
2.1 1000 0.0025
2.1 10000 0.025