Introduction and Specifications
Detailed Specifications 1
1-9
Detailed Specifications
DC Voltage
Absolute Uncertainty, tcal ±5 °C
±(ppm of output +μV)
Stability
Range
90 days 1 year
24 hours, ±1 °C
±(ppm, output +μV)
Resolution μV
Max Burden
[1]
0 to 329.9999 mV
15 + 1 20 + 1 3 + 1
0.1
65 Ω
0 to 3.299999 V 9 + 2 11 + 2 2 + 1.5 1 10 mA
0 to 32.99999 V 10 + 20 12 + 20 2 + 15 10 10 mA
30 V to 329.9999 V 15 + 150 18 + 150 2.5 + 100 100 5 mA
100 V to 1020.000 V 15 + 1500 18 + 1500 3 + 300 1000 5 mA
Auxiliary Output (dual output mode only)
[2]
0 to 329.9999 mV 300 + 350 400 + 350 30 + 100 1 5 mA
0.33 to 3.299999 V 300 + 350 400 + 350 30 + 100 10 5 mA
3.3 to 7 V 300 + 350 400 + 350 30 + 100 100 5 mA
TC Simulate and Measure in Linear 10 μV/°C and 1 mV/°C modes
[3]
0 to 329.9999 mV 40 + 3 50 + 3 5 + 2 0.1
10 Ω
[1] Remote sensing is not provided. Output resistance is <5 mΩ for outputs 0.33 V. The AUX output has an output resistance of
<1 Ω. TC simulation has an output impedance of 10 Ω ±1 Ω.
[2] Two channels of dc voltage output are provided.
[3] TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 v/m.
Noise
Range
Bandwidth 0.1 Hz to 10 Hz p-p ±(ppm
output + floor)
Bandwidth 10 Hz to 10 kHz rms
0 to 329.9999 mV
0 + 1 μV 6 μV
0 to 3.299999 V
0 + 10 μV 60 μV
0 to 32.99999 V
0 + 100 μV 600 μV
30 V to 329.9999 V 10 + 1 mV 20 mV
100 V to 1020.000 V 10 + 5 mV 20 mV
Auxiliary Output (dual output mode only)
[1]
0 to 329.9999 mV
0 + 5 μV 20 μV
0.33 to 3.299999 V
0 + 20 μV 200 μV
3.3 to 7 V
0 + 100 μV 1000 μV
[1] Two channels of dc voltage output are provided.
DC Current
Absolute Uncertainty, tcal ±5 °C
±(ppm of output +μA)
Range
90 days 1 year
Resolution
Max Compliance
Voltage V
Max Inductive
Load mH
0 to 329.999 μA
120 + 0.02 150 + 0.02 1 nA
10
0 to 3.29999 mA
80 + 0.05 100 + 0.05
0.01 μA
10
0 to 32.9999 mA
80 + 0.25 100 + 0.25
0.1 μA
7
0 to 329.999 mA
80 + 2.5 100 + 2.5
1 μA
7
0 to 1.09999 A
160 + 40 200 + 40
10 μA
6
1.1 A to 2.99999 A 300 + 40 380 + 40
10 μA
6
0 to 10.9999 A
(20 A Range)
380 + 500 500 + 500
100 μA
4
11 to 20.5 A
[1]
800 + 750
[2]
1000 + 750
[2]
100 μA
4
400
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1-4. The current may be provided 60-
T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current
in amperes. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5520A is outputting
currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the
allowable "on" time indicated by the formula and Figure B is achieved only after the 5520A is outputting currents <5 A for the "off"
period first.
[2] Floor specification is 1500 μA within 30 seconds of selecting operate. For operating times >30 seconds, the floor specification is
750 μA.