5522A
Operators Manual
10-6
Time Marker Specifications
Time Marker into 50 Ω
5s to 50 ms 20 ms to 100 ns 50 to 20 ns 10 ns 5 to 1 ns
1-Year Absolute Uncertainty at
Cardinal Points, tcal ± 5 °C
±(25 + t x
1000 ppm
[1]
±2.5 ppm ±2.5 ppm ±2.5 ppm ±2.5 ppm
Wave Shape
spike or
square
spike, square, or
20 %-pulse
spike or square
square or
sine
sine
Typical Output Level >1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
Typical Jitter (rms) <10 ppm <1 ppm <1 ppm <1 ppm <1 ppm
Sequence 5-2-1 from 5 s to 1 ns (e.g., 500 ms, 200 ms, 100 ms )
Adjustment Range
[3]
At least ±10 % around each sequence value indicated above.
Amplitude Resolution 4 digits
[1] t is the time in seconds.
[2] Typical rise time of square wave and 20 %-pulse (20 % duty cycle pulse) is <1.5 ns.
[3] Time marker uncertainty is ±50 ppm away from the cardinal points.
Wave Generator Specifications
Wave Generator Characteristics
Square Wave, Sine Wave, and Triangle Wave into 50 Ω or 1 MΩ
Amplitude
Range
into 1 MΩ: 1.8 mV to 55 V p-p
into 50 Ω: 1.8 mV to 2.5 V p-p
1-Year Absolute Uncertainty, tcal
±5 °C, 10 Hz to 10 kHz
±(3 % of p-p output + 100 µV)
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Typical DC Offset Range
0 to ± (40 % of p-p amplitude)
[1]
Frequency
Range 10 Hz to 100 kHz
Resolution 4 or 5 digits depending upon frequency
1-Year Absolute Uncertainty, tcal ± 5 °C ±(25 ppm + 15 mHz)
[1] The dc offset plus the wave signal must not exceed 30 V rms.