EasyManua.ls Logo

Fluke 5522A - Temperature Calibration (Thermocouple)

Fluke 5522A
426 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Detailed Specifications 1
1-17
Temperature Calibration (Thermocouple)
Absolute Uncertainty
Source/Measure
tcal ±5 °C
± °C
[3]
Absolute Uncertainty
Source/Measure
tcal ±5 °C
± °C
[3]
TC
Type
[1]
Range
°C
[2]
90 days 1 year
TC
Type
[1]
Range
°C
[2]
90 days 1 year
600 to 800 0.42 0.44 -200 to -100 0.37 0.37
800 to 1000 0.34 0.34 -100 to 800 0.26 0.26
1000 to 1550 0.30 0.30
L
800 to 900 0.17 0.17
B
1550 to 1820 0.26 0.33 -200 to -100 0.30 0.40
0 to 150 0.23 0.30 -100 to -25 0.17 0.22
150 to 650 0.19 0.26 -25 to 120 0.15 0.19
650 to 1000 0.23 0.31 120 to 410 0.14 0.18
1000 to 1800 0.38 0.50
N
410 to 1300 0.21 0.27
C
1800 to 2316 0.63 0.84 0 to 250 0.48 0.57
-250 to -100 0.38 0.50 250 to 400 0.28 0.35
-100 to -25 0.12 0.16 400 to 1000 0.26 0.33
-25 to 350 0.10 0.14
R
1000 to 1767 0.30 0.40
350 to 650 0.12 0.16 0 to 250 0.47 0.47
E
650 to 1000 0.16 0.21 250 to 1000 0.30 0.36
-210 to -100 0.20 0.27 1000 to 1400 0.28 0.37
-100 to -30 0.12 0.16
S
1400 to 1767 0.34 0.46
-30 to 150 0.10 0.14 -250 to -150 0.48 0.63
150 to 760 0.13 0.17 -150 to 0 0.18 0.24
J
760 to 1200 0.18 0.23 0 to 120 0.12 0.16
-200 to -100 0.25 0.33
T
120 to 400 0.10 0.14
-100 to -25 0.14 0.18 -200 to 0 0.56 0.56
-25 to 120 0.12 0.16
U
0 to 600 0.27 0.27
120 to 1000 0.19 0.26
K
1000 to 1372 0.30 0.40
[1] Temperature standard ITS-90 or IPTS-68 is selectable.
TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
[2] Resolution is 0.01 °C
[3] Does not include thermocouple error

Table of Contents

Related product manuals