Introduction and Specifications
Additional Specifications 1
1-25
AC Current (Non-Sine Wave) (cont.)
Triangle Wave &
Truncated Sine Wave
Range
p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.5
93 to
929.999 mA
[1]
1 to 10 kHz 10 + 2
Six digits
10 to 45 Hz 0.5 + 1.0
45 Hz to 1 kHz 0.5 + 0.5
0.93 to
8.49999 A
1 to 10 kHz 10 + 2
45 to 500 Hz 0.5 + 0.5
8.5 to 57 A
[2]
500 Hz to 1 kHz 1.0 + 1.0
Six digits
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
Square Wave
Range p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
0.047 to
0.65999 mA
[1]
1 to 10 kHz 10 + 2
Six digits
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
0.66 to
6.59999 mA
[1]
1 to 10 kHz 10 + 2
Six digits
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
6.6 to
65.9999 mA
[1]
1 to 10 kHz 10 + 2
Six digits
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.5
66 to
659.999 mA
[1]
1 to 10 kHz 10 + 2
10 to 45 Hz 0.5 + 1.0
45 Hz to 1 kHz 0.5 + 0.5
0.66 to
5.99999 A
[2]
1 to 10 kHz 10 + 2
45 to 500 Hz 0.5 + 0.5
6 to 41 A
[2]
500 Hz to 1 kHz 1.0 + 1.0
Six digits
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
AC Current, Square Wave Characteristics (Typical)
Range LCOMP Risetime Settling Time Overshoot
I <6 A @ 400 Hz off
25 μs 40 μs to 1 % of final value
<10 % for <1 V Compliance
3 A & 20 A Ranges on
100 μs 200 μs to 1 % of final value
<10 % for <1 V Compliance
AC Current, Triangle Wave Characteristics (Typical)
Linearity to 400 Hz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range