SC600 Oscilloscope Calibration Option
Verification Tables 9
9-51
Input Impedance Verification: Resistance
Table 9-19. SC600 Option Input Impedance Verification: Resistance
Nominal Value (Ω) Measured Value (Ω) Deviation (Ω) 1-Year Spec. (Ω)
40 0.04
50 0.05
60 0.06
600000 600
1000000 1000
1500000 1500
Input Impedance Verification: Capacitance
Table 9-20. SC600 Option Input Impedance Verification: Capacitance
Nominal Value (pF) Measured Value (pF) Deviation (pF) 1-Year Spec. (pF)
5 pF 0.75 pF
29 pF 1.95 pF
49 pF 2.95 pF